A Deep-Learning-Based Partial-Volume Correction Method for Quantitative 177Lu SPECT/CT Imaging

delete2024-04-18
delete3
PRE
AI
J
Julian Leube
J
Johan Gustafsson
M
Michael Laßmann
M
Maikol Salas-Ramirez
J
Johannes Tran‐Gia *
DOI:10.2967/jnumed.123.266889delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
With the development of new radiopharmaceutical therapies, quantitative SPECT/CT has progressively emerged as a crucial tool for dosimetry. One major obstacle of SPECT is its poor resolution, which results in blurring of the activity distribution. Especially for small objects, this so-called partial-volume effect limits the accuracy of activity quantification. Numerous methods for partial-volume correction (PVC) have been proposed, but most methods have the disadvantage of assuming a spatially invariant resolution of the imaging system, which does not hold for SPECT. Furthermore, most methods require a segmentation based on anatomic information. Methods: We introduce DL-PVC, a methodology for PVC of 177Lu SPECT/CT imaging using deep learning (DL). Training was based on a dataset of 10,000 random activity distributions placed in extended cardiac-torso body phantoms. Realistic SPECT acquisitions were created using the SIMIND Monte Carlo simulation program. SPECT reconstructions without and with resolution modeling were performed using the CASToR and STIR reconstruction software, respectively. The pairs of ground-truth activity distributions and simulated SPECT images were used for training various U-Nets. Quantitative analysis of the performance of these U-Nets was based on metrics such as the structural similarity index measure or normalized root-mean-square error, but also on volume activity accuracy, a new metric that describes the fraction of voxels in which the determined activity concentration deviates from the true activity concentration by less than a certain margin. On the basis of this analysis, the optimal parameters for normalization, input size, and network architecture were identified. Results: Our simulation-based analysis revealed that DL-PVC (0.95/7.8%/35.8% for structural similarity index measure/normalized root-mean-square error/volume activity accuracy) outperforms SPECT without PVC (0.89/10.4%/12.1%) and after iterative Yang PVC (0.94/8.6%/15.1%). Additionally, we validated DL-PVC on 177Lu SPECT/CT measurements of 3-dimensionally printed phantoms of different geometries. Although DL-PVC showed activity recovery similar to that of the iterative Yang method, no segmentation was required. In addition, DL-PVC was able to correct other image artifacts such as Gibbs ringing, making it clearly superior at the voxel level. Conclusion: In this work, we demonstrate the added value of DL-PVC for quantitative 177Lu SPECT/CT. Our analysis validates the functionality of DL-PVC and paves the way for future deployment on clinical image data.
Keyword:
image processing
SPECT/CT
deep learning
Monte Carlo simulation
dosimetry
partial-volume correction

期刊

Journal of Nuclear Medicine 封面图
Journal of Nuclear Medicine
IF:
9.1
论文数:
6.2K
被引数:
3.0W

机构

L
lund university
学者数:
4.1W
论文数: 3.9W
被引数: 54
U
University of Wurzburg
学者数:
2.5W
论文数: 2.0W
被引数: 2.5W
引用论文

引用论文

Synthesis and antitumor properties of an anthraquinone bisubstituted by the copper chelating peptide Gly-Gly-L-His
err2002-05-01
err0
PREAI
errElisabeth Morier-Teissier; Nezha Boitte; Nicole Helbecque; Jean Luc Bernier; Nicole Pommery; Jean Loup Duvalet; Charles Fournier; Bernard Hecquet; Jean Pierre Catteau; Jean Pierre Henichart
err分享
err收藏
Deep learning with noise-to-noise training for denoising in SPECT myocardial perfusion imaging
err2020-11-23
err39
errOAAI
errLiu, Junchi; Yang, Yongyi; Wernick, Miles N.; Pretorius, P. Hendrik; King, Michael A.
err分享
err收藏
err分享
err收藏
A multicentre and multi-national evaluation of the accuracy of quantitative Lu-177 SPECT/CT imaging performed within the MRTDosimetry project
err2021-07-23
err43
errOAAI
errTran-Gia, Johannes; Denis-Bacelar, Ana M.; Ferreira, Kelley M.; Robinson, Andrew P.; Calvert, Nicholas; Fenwick, Andrew J.; Finocchiaro, Domenico; Fioroni, Federica; Grassi, Elisa; Heetun, Warda; Jewitt, Stephanie J.; Kotzassarlidou, Maria; Ljungberg, Michael; McGowan, Daniel R.; Scott, Nathaniel; Scuffham, James; Gleisner, Katarina Sjogreen; Tipping, Jill; Wevrett, Jill; Lassmann, Michael
err分享
err收藏
CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction
err2018-09-10
err119
errOAAI
errMerlin, Thibaut; Stute, Simon; Benoit, Didier; Bert, Julien; Carlier, Thomas; Comtat, Claude; Filipovic, Marina; Lamare, Frederic; Visvikis, Dimitris
err分享
err收藏
A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology
err2012-10-16
err404
PREAI
errErlandsson, Kjell; Buvat, Irene; Pretorius, P. Hendrik; Thomas, Benjamin A.; Hutton, Brian F.
err分享
err收藏
The effect of ligand amount, affinity and internalization on PSMA-targeted imaging and therapy: A simulation study using a PBPK model
err2019-12-27
err38
errOAAI
errBegum, Nusrat J.; Glatting, Gerhard; Wester, Hans-Juergen; Eiber, Matthias; Beer, Ambros J.; Kletting, Peter
err分享
err收藏
学者 查看更多内容