A digitally reconstructed radiograph algorithm calculated from first principles

delete2012-12-14
delete22
delete
OA
AI
D
David Staub *
M
Martin J. Murphy
DOI:10.1118/1.4769413delete
delete原文链接
delete分享
delete收藏
查看原文
摘要

摘要

En 中文
Purpose: To develop an algorithm for computing realistic digitally reconstructed radiographs (DRRs) that match real cone-beam CT (CBCT) projections with no artificial adjustments. Methods: The authors used measured attenuation data from cone-beam CT projection radiographs of different materials to obtain a function to convert CT number to linear attenuation coefficient (LAC). The effects of scatter, beam hardening, and veiling glare were first removed from the attenuation data. Using this conversion function the authors calculated the line integral of LAC through a CT along rays connecting the radiation source and detector pixels with a ray-tracing algorithm, producing raw DRRs. The effects of scatter, beam hardening, and veiling glare were then included in the DRRs through postprocessing. Results: The authors compared actual CBCT projections to DRRs produced with all corrections (scatter, beam hardening, and veiling glare) and to uncorrected DRRs. Algorithm accuracy was assessed through visual comparison of projections and DRRs, pixel intensity comparisons, intensity histogram comparisons, and correlation plots of DRR-to-projection pixel intensities. In general, the fully corrected algorithm provided a small but nontrivial improvement in accuracy over the uncorrected algorithm. The authors also investigated both measurement-and computation-based methods for determining the beam hardening correction, and found the computation-based method to be superior, as it accounted for nonuniform bowtie filter thickness. The authors benchmarked the algorithm for speed and found that it produced DRRs in about 0.35 s for full detector and CT resolution at a ray step-size of 0.5 mm. Conclusions: The authors have demonstrated a DRR algorithm calculated from first principles that accounts for scatter, beam hardening, and veiling glare in order to produce accurate DRRs. The algorithm is computationally efficient, making it a good candidate for iterative CT reconstruction techniques that require a data fidelity term based on the matching of DRRs and projections. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4769413]
Keyword:
digitally reconstructed radiograph
DRR
scatter
beam hardening
veiling glare
AI总结

AI总结

对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。

期刊

Medical Physics 封面图
Medical Physics
IF:
3.2
论文数:
3.7W
被引数:
3.2W

机构

V
Virginia Commonwealth University
学者数:
2.2W
论文数: 1.8W
被引数: 1.9W
引用论文

引用论文

Signatures from clustered airguns
err1992-08-01
err0
PREAI
errS. Strandenes; S. Vaage
err分享
err收藏
Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT
err2010-09-28
err35
errOAAI
errLazos, Dimitrios; Williamson, Jeffrey F.
err分享
err收藏
err分享
err收藏
Post-stack reverse-time migration using a finite difference method based on triangular grids
err2008-07-20
err0
PREAI
errShujuan Guo; Zhenchun Li; Xiaodong Sun; Yueming Ye; Houhua Teng; Fang Li
err分享
err收藏
The Function of Powder Downs in Herons
err1920-09-01
err0
PREAI
errAlexander Wetmore
err分享
err收藏
Reconstruction of a cone-beam CT image via forward iterative projection matching
err2010-11-10
err34
errOAAI
errBrock, R. Scott; Docef, Alen; Murphy, Martin J.
err分享
err收藏
学者 查看更多内容