A digitally reconstructed radiograph algorithm calculated from first principles
DOI:10.1118/1.4769413.png)
摘要
En 中文
Purpose: To develop an algorithm for computing realistic digitally reconstructed radiographs (DRRs) that match real cone-beam CT (CBCT) projections with no artificial adjustments. Methods: The authors used measured attenuation data from cone-beam CT projection radiographs of different materials to obtain a function to convert CT number to linear attenuation coefficient (LAC). The effects of scatter, beam hardening, and veiling glare were first removed from the attenuation data. Using this conversion function the authors calculated the line integral of LAC through a CT along rays connecting the radiation source and detector pixels with a ray-tracing algorithm, producing raw DRRs. The effects of scatter, beam hardening, and veiling glare were then included in the DRRs through postprocessing. Results: The authors compared actual CBCT projections to DRRs produced with all corrections (scatter, beam hardening, and veiling glare) and to uncorrected DRRs. Algorithm accuracy was assessed through visual comparison of projections and DRRs, pixel intensity comparisons, intensity histogram comparisons, and correlation plots of DRR-to-projection pixel intensities. In general, the fully corrected algorithm provided a small but nontrivial improvement in accuracy over the uncorrected algorithm. The authors also investigated both measurement-and computation-based methods for determining the beam hardening correction, and found the computation-based method to be superior, as it accounted for nonuniform bowtie filter thickness. The authors benchmarked the algorithm for speed and found that it produced DRRs in about 0.35 s for full detector and CT resolution at a ray step-size of 0.5 mm. Conclusions: The authors have demonstrated a DRR algorithm calculated from first principles that accounts for scatter, beam hardening, and veiling glare in order to produce accurate DRRs. The algorithm is computationally efficient, making it a good candidate for iterative CT reconstruction techniques that require a data fidelity term based on the matching of DRRs and projections. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4769413]
Keyword:
digitally reconstructed radiograph
DRR
scatter
beam hardening
veiling glare
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
3.2
论文数:
3.7W
被引数:
3.2W
机构
引用论文
Reconstruction of a cone-beam CT image via forward iterative projection matching
MEDICAL PHYSICS
IF3.2