En 中文
The design of a simple and inexpensive battery powered lamp controller for microscope lamps commonly used in patch clamp and other biophysical studies is described. The circuit linearizes the normally exponential voltage-intensity curve of standard 12-V lamps by using an operational amplifier to provide feedback to the drain of a power MOSFET. When the unit is turned off, a charging circuit recharges the battery at a constant potential until the battery voltage reaches a preset level. The charging voltage is then reduced to balance the self-discharge rate of the battery.