A Novel Method for Quantifying Scanner Instability in fMRI
DOI:10.1002/mrm.22691.png)
摘要
En 中文
A method was developed to quantify the effect of scanner instability on functional MRI data by comparing the instability noise to endogenous noise present when scanning a human. The instability noise was computed from agar phantom data collected with two flip angles, allowing for a separation of the instability from the background noise. This method was used on human data collected at four 3 T scanners, allowing the physiological noise level to be extracted from the data. In a well-operating scanner, the instability noise is generally less than 10% of physiological noise in white matter and only about 2% of physiological noise in cortex. This indicates that instability in a well-operating scanner adds very little noise to functional MRI results. This new method allows researchers to make informed decisions about the maximum instability level a scanner can have before it is taken off line for maintenance or rejected from a multisite consortium. This method also provides information about the background noise, which is generally larger in magnitude than the instability noise. Magn Reson Med 65:1053-1061, 2011. (C) 2010 Wiley-Liss, Inc.
Keyword:
quality assurance
fMRI
multisite
scanner instability