A prototype MR insertable brain PET using tileable GAPD arrays

delete2013-03-13
delete91
PRE
AI
K
Key Jo Hong
Y
Yong Choi *
J
Jin Ho Jung
J
Jihoon Kang
W
Wei Hu
H
Hyun Keong Lim
Y
Yoonsuk Huh
S
Sangsu Kim
J
Ji Woong Jung
K
Kyu Bom Kim
M
Myung Sung Song
H
HyunWook Park
DOI:10.1118/1.4793754delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
Purpose: The aim of this study was to develop a prototype magnetic resonance (MR)-compatible positron emission tomography (PET) that can be inserted into a MR imager and that allows simultaneous PET and MR imaging of the human brain. This paper reports the initial results of the authors' prototype brain PET system operating within a 3-T magnetic resonance imaging (MRI) system using newly developed Geiger-mode avalanche photodiode (GAPD)-based PET detectors, long flexible flat cables, position decoder circuit with high multiplexing ratio, and digital signal processing with field programmable gate array-based analog to digital converter boards. Methods: A brain PET with 72 detector modules arranged in a ring was constructed and mounted in a 3-T MRI. Each PET module was composed of cerium-doped lutetium yttrium orthosilicate (LYSO) crystals coupled to a tileable GAPD. The GAPD output charge signals were transferred to preamplifiers using 3 m long flat cables. The LYSO and GAPD were located inside the MR bore and all electronics were positioned outside the MR bore. The PET detector performance was investigated both outside and inside the MRI, and MR image quality was evaluated with and without the PET system. Results: The performance of the PET detector when operated inside the MRI during MR image acquisition showed no significant change in energy resolution and count rates, except for a slight degradation in timing resolution with an increase from 4.2 to 4.6 ns. Simultaneous PET/MR images of a hot-rod and Hoffman brain phantom were acquired in a 3-T MRI. Rods down to a diameter of 3.5 mm were resolved in the hot-rod PET image. The activity distribution patterns between the white and gray matter in the Hoffman brain phantom were well imaged. The hot-rod and Hoffman brain phantoms on the simultaneously acquired MR images obtained with standard sequences were observed without any noticeable artifacts, although MR image quality requires some improvement. Conclusions: These results demonstrate that the simultaneous acquisition of PET and MR images is feasible using the MR insertable PET developed in this study. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4793754]
Keyword:
PET
brain imaging
MRI
GAPD arrays
LYSO
position decoder circuit

期刊

Medical Physics 封面图
Medical Physics
IF:
3.2
论文数:
3.7W
被引数:
3.2W

机构

S
Sogang University
学者数:
4.6K
论文数: 4.4K
被引数: 4.0K
引用论文

引用论文

err
IF0
err
err0
PREAI
err
err分享
err收藏
Lateral Force during Curve Negotiation of Forced Steering Bogies
err2003-01-01
err0
errOAAI
errEisaku SATO; Hideyuki KOBAYASHI; Kazuhiko TEZUKA; Isao OKAMOTO; Hirohiko KAKINUMA; Toshiharu TAMAOKI
err分享
err收藏
Virtual prototyping for customized product development
err1998-12-01
err0
errOAAI
errMitchell M. Tseng; Jianxin Jiao; Chuan‐Jun Su
err分享
err收藏
Up-regulation and overproduction of DVL-1, the human counterpart of the Drosophila dishevelled gene, in cervical squamous cell carcinoma
err2003-09-01
err0
PREAI
errKeiko Okino; Hisaki Nagai; Mitsuko Hatta; Takemitsu Nagahata; Koichi Yoneyama; Yujiro Ohta; Enjing Jin; Oichi Kawanami; Tsutomu Araki; Mitsuru Emi
err分享
err收藏
学者 查看更多内容