arrow
返回

Absorption detection using optical waveguide cavities

delete2010-05-01
delete0
delete
OA
AI
DOI:10.1139/v10-006delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
Cavity ring-down spectroscopy is a spectroscopic method that uses a high quality optical cavity to amplify the optical loss due to the light absorption by a sample. In this presentation we highlight two applications of phase-shift cavity ring-down spectroscopy that are suited for absorption measurements in the condensed phase and make use of waveguide cavities. In the first application, a fiber loop is used as an optical cavity and the sample is introduced in a gap in the loop to allow absorption measurements of nanoliters of solution at the micromolar level. A second application involves silica microspheres as high finesse cavities. Information on the refractive index and absorption of a thin film of ethylene diamine on the surface of the microresonator is obtained simultaneously by the measurements of the wavelength shift of the cavity mode spectrum and the change in optical decay time, respectively.

期刊

暂无期刊信息

机构

暂无机构信息
引用论文

引用论文

暂无论文信息