Brachytherapy needle deflection evaluation and correction
DOI:10.1118/1.1871372.png)
摘要
En 中文
In prostate brachytherapy, an 18-gauge needle is used to implant radioactive seeds. This thin needle can be deflected from the preplanned trajectory in the prostate, potentially resulting in a suboptimum dose pattern and at times requiring repeated needle insertion to achieve optimal dosimetry. In this paper, we report on the evaluation of brachytherapy needle deflection and bending in test phantoms and two approaches to overcome the problem. First we tested the relationship between needle deflection and insertion depth as well as whether needle bending occurred. Targeting accuracy was tested by inserting a brachytherapy needle to target 16 points in chicken tissue phantoms. By implanting dummy seeds into chicken tissue phantoms under 3D ultrasound guidance, the overall accuracy of seed implantation was determined. We evaluated methods to overcome brachytherapy needle deflection with three different insertion methods: constant orientation, constant rotation, and orientation reversal at half of the insertion depth. Our results showed that needle deflection is linear with needle insertion depth, and that no noticeable bending occurs with needle insertion into the tissue and agar phantoms. A 3D principal component analysis was performed to obtain the population distribution of needle tip and seed position relative to the target positions. Our results showed that with the constant orientation insertion method, the mean needle targeting error was 2.8 mm and the mean seed implantation error was 2.9 mm. Using the constant rotation and orientation reversal at half insertion depth methods, the deflection error was reduced. The mean needle targeting errors were 0.8 and 1.2 mm for the constant rotation and orientation reversal methods, respectively, and the seed implantation errors were 0.9 and 1.5 mm for constant rotation insertion and orientation reversal methods, respectively. (c) 2005 American Association of Physicists in Medicine.
Keyword:
needle divergence
prostate brachytherapy
3D ultrasound-guided therapy
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
3.2
论文数:
3.7W
被引数:
3.2W
机构
暂无机构信息
引用论文
Comparison of core needle breast biopsy techniques: Freehand versus three-dimensional US guidance
ACADEMIC RADIOLOGY
IF3.9
Treatment profiles and clinical outcomes of COVID-19 patients at private hospital in Jakarta
PLOS ONE
IF0
Robot-assisted 3D-TRUS guided prostate brachytherapy: System integration and validation
MEDICAL PHYSICS
IF3.2
Dosimetric effects of needle divergence in prostate seed implant using 125I and 103Pd radioactive seeds
MEDICAL PHYSICS
IF3.2