Calibration procedures for microwave vector comparators
DOI:10.1016/j.sna.2021.112955.png)
摘要
En 中文
This paper presents two calibration procedures for microwave vector comparators. The first is intended for architectures composed of a quadrature demodulator. It assumes that the output voltages of this component are expressed as cosine and sine functions with different amplitudes and DC offsets. Tests with a prototype using an ADL5380 have shown amplitude and phase errors less than 0.1 dB and 4.3 degrees, respectively. The second calibration procedure is developed for comparators comprised of gain and phase detectors. A prototype with two AD8302 was designed to provide 360 degrees phase coverage, since the AD8302 is limited to 180 degrees phase detection. As the developed comparator couples the input signals internally, we introduce an efficient way of determining the corresponding coupling coefficients to ensure high accuracy. Tests with the prototype have shown amplitude and phase errors below 0.7 dB and 2 degrees, respectively. Finally, experiments conducted in an anechoic chamber with a phased array and a direction finding system using the two prototypes demonstrate the efficacy of the calibration procedures. (C) 2021 Elsevier B.V. All rights reserved.
Keyword:
Direction finding system
Gain and phase detector
Phased array
Quadrature demodulator
Vector comparator
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
4.9
论文数:
1.5W
被引数:
3.3W
机构
暂无机构信息
引用论文
Phase detection using AD8302 evaluation board in the superheterodyne microwave interferometer for line average plasma electron density measurements
MEASUREMENT
IF5.6
Kinetic and Equilibrium Studies on the Adsorption of Crystal Violet Dye using Kaolin as an Adsorbent