返回
Capillary-based displacement flow immunosensor
DOI:10.1021/ac961037y.png)
摘要
En 中文
We report on the preparation of a novel capillary flow immunosensor for 2,4,6-trinitrotoluene (TNT). Anti-TNT antibody is immobilized onto the silanized inner walls of an 800-mu m-i.d. glass capillary using a heterobifunctional crosslinker, followed by saturating the capillary with fluorophore-labeled antigen. To perform the assay, an aliquot of TNT solution is injected into the capillary. TNT competes for the binding pocket of the fluorophore-labeled antigen, which is detected downstream, We have demonstrated that the sensitivity of the capillary immunosensor is improved by at least 2 orders of magnitude compared to that of conventional packed-column;or membrane-based displacement flow immunosensors. The limit of detection (S/N > 3) for TNT using the capillary immunosensor is 15 pg/mL (7 fmol), and the linear dynamic range is 1-250 ng/mL. The entire assay can be performed in less than 3 min. During 15 injections of 100 ng/mL TNT solution, the signal reproducibility (RSD) was maintained to within 4%.
Keyword:
ENZYME-IMMUNOASSAY
2,4,6-TRINITROTOLUENE
WATER
TNT
EXPLOSIVES
TRINITROTOLUENE
ANTIBODIES
BIOSENSOR
ATRAZINE
ELISA
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
6.7
论文数:
4.7W
被引数:
15.9W
机构
暂无机构信息

