arrow
返回

Cost-efficient unsupervised sample selection for multivariate calibration

delete2021-08-01
delete11
delete
OA
AI
V
Valeria Fonseca Díaz *
B
Bart De Ketelaere
B
Ben Aernouts
W
Wouter Saeys
DOI:10.1016/j.chemolab.2021.104352delete
delete原文链接
delete分享
delete收藏
查看原文
摘要

摘要

En 中文
Indirect quantification of chemical composition through spectral measurements requires the establishment of multivariate calibration models. The reference analyses on the calibration samples typically form a major cost factor in the establishment of these multivariate models. Therefore, the aim of this study was to select the most informative calibration samples in an unsupervised way based on the spectral measurements. To this end, guidelines to address this challenge in PLSR model building have been developed. The recommendations include calculating a sample size that surpasses the model complexity by a factor of 12, performing the selection in the PCA score space spanned by a sufficiently large number of principal components and using methods such as Kennard-Stone, Puchwein, Clustering or D-optimal designs. We provide the data and methodology used in the present study for future use.
Keyword:
Sample selection
Unsupervised learning
Multivariate calibration
PLS regression
VC dimension
NIR