Cuffless Blood Pressure Estimation Model Using Clustering Techniques
DOI:10.1109/JSEN.2023.3327683.png)
摘要
En 中文
This study introduces a clustered algorithm for estimating blood pressure (BP) using electrocardiographic and photoplethysmographic data obtained from 1675 subjects, corresponding to a total of 1 051 462 cardiac cycles. These data were sourced from the publicly accessible biosignal database, VitalDB. To enable the application of pulse-wave analysis (PWA) on the recorded signals, the study extracted 29 morphological features from these signals and employed the K-means clustering algorithm to categorize them into six distinct clusters. Each of these clusters was used to train and optimize three separate BP estimation algorithms. Notably, the Shapley additive explanations analysis identified arterial stiffness and peripheral resistance as the most influential factors contributing to the clustering process. Experimental results reveal that the most accurate BP estimation model achieves deviations of 0.043 +/- 6.43 mmHg for systolic BP and 0.052 +/- 3.65 mmHg for diastolic BP (DBP). Remarkably, the adoption of clustering leads to notable enhancements in accuracy across key metrics, including mean error (ME), standard deviation, and mean absolute error estimates. Importantly, these improvements align with the rigorous criteria established by three international BP standards: The Association for the Advancement of Medical Instrumentation/ISO, the British Hypertension Society (BHS), and the Institution of Electrical and Electronic Engineers. This study underscores the potential for clinical adoption of the clustering-based, cuffless BP estimation algorithm, which accommodates individual variations without necessitating a pretraining process for the subjects under investigation.
Keyword:
Clustering
cuffless blood pressure (BP) estimation
hypertension
pulse-wave analysis (PWA)
期刊
IF:
4.5
论文数:
2.2W
被引数:
7.3W
机构
引用论文
Vital Recorder-a free research tool for automatic recording of high-resolution time-synchronised physiological data from multiple anaesthesia devices
SCIENTIFIC REPORTS
IF3.9
Estimation and Validation of Arterial Blood Pressure Using Photoplethysmogram Morphology Features in Conjunction With Pulse Arrival Time in Large Open Databases在大型开放数据库中使用光学体积描记图形态特征结合脉搏到达时间对动脉血压的估计和验证
Towards accurate estimation of cuffless and continuous blood pressure using multi-order derivative and multivariate photoplethysmogram features使用多阶导数和多变量光电体积描记图特征准确估计无袖带和连续血压