Evaluating Machine Learning-Based MRI Reconstruction Using Digital Image Quality Phantoms
DOI:10.3390/bioengineering11060614.png)
摘要
En 中文
Quantitative and objective evaluation tools are essential for assessing the performance of machine learning (ML)-based magnetic resonance imaging (MRI) reconstruction methods. However, the commonly used fidelity metrics, such as mean squared error (MSE), structural similarity (SSIM), and peak signal-to-noise ratio (PSNR), often fail to capture fundamental and clinically relevant MR image quality aspects. To address this, we propose evaluation of ML-based MRI reconstruction using digital image quality phantoms and automated evaluation methods. Our phantoms are based upon the American College of Radiology (ACR) large physical phantom but created in k-space to simulate their MR images, and they can vary in object size, signal-to-noise ratio, resolution, and image contrast. Our evaluation pipeline incorporates evaluation metrics of geometric accuracy, intensity uniformity, percentage ghosting, sharpness, signal-to-noise ratio, resolution, and low-contrast detectability. We demonstrate the utility of our proposed pipeline by assessing an example ML-based reconstruction model across various training and testing scenarios. The performance results indicate that training data acquired with a lower undersampling factor and coils of larger anatomical coverage yield a better performing model. The comprehensive and standardized pipeline introduced in this study can help to facilitate a better understanding of the performance and guide future development and advancement of ML-based reconstruction algorithms.
Keyword:
machine learning
MRI reconstruction
automated image quality evaluation
digital image quality phantom
image resolution
low-contrast detectability
digital reference object
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
B
IF:
3.7
论文数:
5.9K
被引数:
1.3W
机构
暂无机构信息
引用论文
Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms
MEDICAL PHYSICS
IF3.2
Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain
MEDICAL PHYSICS
IF3.2
Boosting the signal-to-noise of low-field MRI with deep learning image reconstruction
SCIENTIFIC REPORTS
IF3.9