Explosives detection in soil using a field-portable continuous flow immunosensor
DOI:10.1016/S0304-3894(00)00327-7.png)
摘要
En 中文
A field method for quantitative analysis of explosives in contaminated soil samples is described. The method is based on a displacement immunoassay performed in a commercial instrument, the FAST 2000, engineered by Research International Inc. The method can be used on-sire to measure 2,4,6-trinitrotoluene (TNT) and hexahydro-1,3,5-trinitro-1,3,5-triazine (RDX) within 5 min. For this study, replicate analyses were performed on soil extracts prepared from each field sample as well as appropriate controls, blanks, and laboratory standards. Statistical analyses were done to assess accuracy, bias, and predictability of the method. The results demonstrated that the immunosensor could be used effectively to screen environmental samples for the presence or absence of explosives. In most samples, the method also provided quantitative values that were in good agreement with standard laboratory analyses using HPLC. A limited number of sample matrices interfered with the immunoassay and produced results that varied significantly from the laboratory data. In each case, the compounds causing the problem have been identified and efforts are being made to minimize these matrix interferences in future field evaluations. (C) 2001 Elsevier Science B.V. All rights reserved.
Keyword:
explosives
soil
immunosensor
field methods