Galvanic Impulse Wireless Communication for Biomedical Implants
DOI:10.1109/ACCESS.2021.3064206.png)
摘要
En 中文
Wireless intra-body communication is a promising approach for providing efficient and secure connectivity for medical implants. The low power consumption of the electronics and the conductivity of the biological tissues facilitate the system implementation, which makes the technique more power-efficient than the traditional radio frequency wireless systems. The galvanic intra-body communication uses the electrical current for signal transmission in the conductive medium of the biological tissues. In this paper, we propose an ultra-low-power communication approach by implementing a galvanic impulse method for communication between an implant and an on-body device. The communication system is designed, manufactured with off-the-shelf electronic components, and measured in the phantom and in-vivo animal experiment. The implant power consumption is 45 mu W for the data rate of 64 kbps with a bit error percentage below 0.5% for the implant depth of 14 cm. The design supports long-lasting battery-powered implant sensory and communication system.
Keyword:
Galvanic coupling
intra-body communication
biomedical implants
biomedical electronics
ultra-wideband