High-Current CT Calibration Using a Sampling Current Ratio Bridge

delete2013-06-01
delete13
PRE
AI
G
Gert Rietveld *
L
Leendert Jol
H
Helko E. van den Brom
E
E. So
DOI:10.1109/TIM.2013.2253975delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
A measurement setup is developed for the accurate ratio measurement of ac current transformers (CTs) for primary currents up to 5 kA. A unique property of the system is the use of high-quality digitizers for sampling of the secondary current signals and step-down transformers with different current ratios. This allows for accurate comparison of CTs even when their nominal ratios are not equal. The calibration results of the key components in the setup show that the sampling current ratio bridge has a ratio uncertainty of better than 2 mu A/A in magnitude and 0.8 mu rad in phase for CT-under-test current ratios that do not differ by more than a factor of 5 from one of the ratios in the reference CT. The linearity of the bridge for input currents changing between 1% and 120% of nominal input current is better than 4 mu A/A in magnitude and 0.5 mu rad in phase.
Keyword:
Alternating current
current measurement
current ratio
current transformers
digital sampling
phase displacement
phase error
ratio error
AI总结

AI总结

对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。

期刊

IEEE Transactions on Instrumentation and Measurement 封面图
IEEE Transactions on Instrumentation and Measurement
IF:
5.9
论文数:
2.0W
被引数:
5.8W

机构

V
VSL Dutch Metrology Institute
学者数:
128
论文数: 99
被引数: 94
N
National Research Council Canada
学者数:
7.9K
论文数: 7.9K
被引数: 6.8K
引用论文

引用论文

err
IF0
err
err0
PREAI
err
err分享
err收藏
err分享
err收藏
Possible association between dysnatremias and mortality during hospitalization in patients undergoing acute hemodialysis: analysis from a Peruvian retrospective cohort
err2019-12-01
err0
errOAAI
errEdward Mezones-Holguin; Roberto Niño-Garcia; Percy Herrera-Añazco; Álvaro Taype-Rondan; Josmel Pacheco-Mendoza; Adrian V. Hernandez
err分享
err收藏
没有更多内容