HIGH-RESOLUTION MR IMAGING USING LOOP-GAP RESONATORS - WORK IN PROGRESS

delete1986-01-01
delete35
PRE
AI
J
J. Bruce Kneeland
A
Andrzej Jesmanowicz
W
Wojciech Froncisz
T
Thomas M. Grist
J
J S Hyde
DOI:10.1148/radiology.158.1.3940388delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
Representative results using loop-gap resonator (LGR) configurations for local high-resolution MR imaging (0.6 mm .times. 0.6 mm pixel size with 3-mm section thickness) are presented. Four geometries were employed: flat single loops, coplanar (planar pair) loops, coaxial (axial pair) loops, and symmetrically angled (butterfly pair) loops. The LGRs were used as receiver coils with the body coil as transmitter. Butterfly pairs and planar pairs are intrinsically decoupled from a transmitted field of arbitrary orientation, whereas the single loops and axial pairs are decoupled from a field that is linearly polarized and orthogonal to its axis. The coils were evaluated using computer-generated and experimentally determined field distributions and by the imaging of human subjects on a 1.5-T MR system. Matching the region of sensitivity of the coil to the region of anatomic interest is emphasized.