arrow
返回

Low cost capillary flow meter

delete2009-06-01
delete17
PRE
AI
M
Mateusz Kudasik
N
Norbert Skoczylas *
J
Jacek Sobczyk
DOI:10.1016/j.sna.2009.03.023delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
This article introduces a project and some specifications of the capillary flow meter. The functioning of the presented flow meter is based on the measurement of pressure drop caused by the gas flowing through the capillary. The article includes in detail a description of the calibration process of the above-mentioned measuring device and presents its basic specifications and capabilities. The measurement results are barometric and temperature compensated. The upper range boundary of the presented device amounts to 45 mu mol/s and the measuring error accounts for 0.3% of the full measuring range of the meter. The costs of material used for constructing the prototype of the capillary flow meter did not exceed US$ 30. (C) 2009 Elsevier BY. All rights reserved.
Keyword:
Flow meter
Capillary
Differential pressure
Flow meter calibration
Capillary calibration
AI总结

AI总结

对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。

期刊

Sensors and Actuators A-Physical 封面图
Sensors and Actuators A-Physical
IF:
4.9
论文数:
1.5W
被引数:
3.3W

机构

P
Polish Academy of Sciences
学者数:
3.0W
论文数: 3.1W
被引数: 3.1W
引用论文

引用论文