Motion compensation for aortic valves using partial angle CT reconstructions
DOI:10.1002/mp.15379.png)
摘要
En 中文
Purpose A motion compensation method that is aimed at correcting motion artifacts of cardiac valves is proposed. The primary focus is the aortic valve. Methods The method is based around partial angle reconstructions and a cost function including the image entropy. A motion model is applied to approximate the cardiac motion in the temporal and spatial domain. Based on characteristic values for velocities and strain during cardiac motion, penalties for the velocity and spatial derivatives are introduced to maintain anatomically realistic motion vector fields and avoid distortions. The model addresses global elastic deformation, but not the finer and more complicated motion of the valve leaflets. Results The method is verified based on clinical data. Image quality was improved for most artifact-impaired reconstructions. An image quality study with Likert scoring of the motion artifact severity on a scale from 1 (highest image quality) to 5 (lowest image quality/extreme artifact presence) was performed. The biggest improvements after applying motion compensation were achieved for strongly artifact-impaired initial images scoring 4 and 5, resulting in an average change of the scores by -0.59 +/- 0.06$-0.59\pm 0.06$ and -1.33 +/- 0.03$-1.33\pm 0.03$, respectively. In the case of artifact-free images, a chance to introduce blurring was observed and their average score was raised by 0.42 +/- 0.03. Conclusion Motion artifacts were consistently removed and image quality improved.
Keyword:
cardiac CT
cardiac valve
motion compensation
期刊
IF:
3.2
论文数:
3.7W
被引数:
3.2W
机构
引用论文
Nonrigid registration-based coronary artery motion correction for cardiac computed tomography
MEDICAL PHYSICS
IF3.2
Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction
MEDICAL PHYSICS
IF3.2
ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart来自心脏的亚秒多层螺旋ct扫描的ECG相关图像重建
MEDICAL PHYSICS
IF3.2
Direct cone-beam cardiac reconstruction algorithm with cardiac banding artifact correction
MEDICAL PHYSICS
IF3.2
Improving best-phase image quality in cardiac CT by motion correction with MAM optimization通过MAM优化运动校正提高心脏CT最佳相位图像质量
MEDICAL PHYSICS
IF3.2