Photon Starvation Artifact Reduction by Shift-Variant Processing

delete2022-01-01
delete5
delete
OA
AI
G
Gengsheng L. Zeng *
DOI:10.1109/ACCESS.2022.3142775delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
The x-ray computed tomography (CT) images with low dose are noisy and may contain photon starvation artifacts. The artifacts are location and direction dependent. Therefore, the common shift-invariant denoising filters do not work well. The state-of-the-art methods to process the low-dose CT images are image reconstruction based; they require the raw projection data. In many situations, the raw CT projections are not accessible. This paper suggests a method to denoise the low-dose CT image using the pseudo projections generated by the application of a forward projector on the low-dose CT image. The feasibility of the proposed method is demonstrated by real clinical data.
Keyword:
Nonlinear filters
Maximum likelihood detection
Image reconstruction
Computed tomography
X-ray imaging
Noise measurement
Photonics
Image processing
image reconstruction
biomedical imaging
computed tomography
filters

期刊

IEEE Access 封面图
IEEE Access
IF:
3.6
论文数:
9.8W
被引数:
29.4W

机构

U
Utah System of Higher Education
学者数:
4.6W
论文数: 4.0W
被引数: 161
引用论文

引用论文

err分享
err收藏
Domain Progressive 3D Residual Convolution Network to Improve Low-Dose CT Imaging域渐进三维残差卷积网络改善低剂量CT成像
err2019-12-01
err176
PREAI
errYin, Xiangrui; Zhao, Qianlong; Liu, Jin; Yang, Wei; Yang, Jian; Quan, Guotao; Chen, Yang; Shu, Huazhong; Luo, Limin; Coatrieux, Jean-Louis
err分享
err收藏
Image detail enhancement with spatially guided filters
err2016-03-01
err26
PREAI
errHao, Shijie; Pan, Daru; Guo, Yanrong; Hong, Richang; Wang, Meng
err分享
err收藏
学者 查看更多内容