Background: The intrinsic (muscular) patient effort driving inspiration in non-invasive ventilation modes, such as continuous positive airway pressure (CPAP) therapy, has not been identified from non-invasive data. Current CPAP settings are based on clinical judgment and assessment of symptoms of respiratory distress. Non-optimal settings, including too much positive end expiratory pressure (PEEP) can cause unintended lung injury and ventilator unloading, where patient effort drops and the CPAP device enables too much work being imposed on the injured lung. Currently, there is no non-invasive means of quantifying or identifying these effects. Methods: A novel model-based method of ascertaining intrinsic patient work of breathing (WOB) in CPAP is developed based on linear single compartment and 2nd order b-spline models previously used in invasive ventilation modes. Results are compared to current clinical indications, such as total Imposed WOB from the CPAP device and beak length, the latter of which is the clinical metric used to indicate alveolar overdistension. Intrinsic and Imposed WOB are compared. The hypothesis is that ventilator unloading can be assessed as a decrease in Intrinsic WOB relative to Imposed WOB, as PEEP and associated ventilator unloading rise. This hypothesis is tested using 14 subjects from a CPAP trial of several breathing rates at two PEEP levels. Results: The ratio of Intrinsic to Imposed WOB, normalised per unit tidal volume, decreased with increasing PEEP (4-7 cm H2O), capturing the expected trend of ventilator unloading. Ventilator unloading was observed across all breathing rates. Beak length measurements showed no conclusive evidence of capturing overdistension at higher PEEP or ventilator unloading. Conclusions: Patient Intrinsic WOB in CPAP was non-invasively quantified using model-based methods, based on pressure and flow measurements. The ratio of Intrinsic to Imposed WOB per unit tidal volume clearly and consistently showed ventilator unloading across all patients and breathing rates, with Intrinsic WOB decreasing with increasing PEEP. This trend was not observed in the current clinical metric of beak length. Non-invasively quantifying Intrinsic WOB and ventilator unloading is the critical first step to objectively optimising clinical CPAP settings, patient care, and outcomes.
Keyword:
CPAP
PEEP
Patient effort
Work of breathing
Ventilator unloading
Respiratory mechanics
Respiratory modelling
期刊
IF:
6.3
论文数:
8.3K
被引数:
3.3W
机构
引用论文
High levels of PEEP may improve survival in acute respiratory distress syndrome: A meta-analysis
RESPIRATORY MEDICINE
IF3.1
Titration and Implementation of Neurally Adjusted Ventilatory Assist in Critically III Patients
CHEST
IF8.6
Characterisation of aqueous waste produced during the clandestine production of amphetamine following the Leuckart route utilising solid‐phase extraction gas chromatography–mass spectrometry and capillary electrophoresis with contactless conductivity detection使用固相萃取气相色谱-质谱法和毛细管电泳和非接触式电导检测,按照Leuckart路线秘密生产苯丙胺过程中产生的含水废物的表征
Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome - A randomized controlled trial急性肺损伤和急性呼吸窘迫综合征成人的呼气末正压设置-一项随机对照试验