Scatter correction method for x-ray CT using primary modulation: Phantom studies

delete2010-01-28
delete85
delete
OA
AI
高河伟 封面图
高河伟 (Hewei Gao) *
R
Rebecca Fahrig
N
N. Robert Bennett
M
Mingshan Sun
朱磊 封面图
朱磊 (Lei Zhu)
DOI:10.1118/1.3298014delete
delete原文链接
delete原文求助
delete分享
delete收藏
摘要

摘要

En 中文
Methods: The authors first analyze the potential errors of the estimated scatter in the primary modulation method. On two tabletop CT systems, the method is investigated using three phantoms: A Catphan(C)600 phantom, an anthropomorphic chest phantom, and the Catphan(C)600 phantom with two annuli. Two different primary modulators are also designed to show the impact of the modulator parameters on the scatter correction efficiency. The first is an aluminum modulator with a weak modulation and a low modulation frequency, and the second is a copper modulator with a strong modulation and a high modulation frequency. Results: On the Catphan(C)600 phantom in the first study, the method reduces the error of the CT number in the selected regions of interest (ROIs) from 371.4 to 21.9 Hounsfield units (HU); the contrast to noise ratio also increases from 10.9 to 19.2. On the anthropomorphic chest phantom in the second study, which represents a more difficult case due to the high scatter signals and object heterogeneity, the method reduces the error of the CT number from 327 to 19 HU in the selected ROIs and from 31.4% to 5.7% on the overall average. The third study is to investigate the impact of object size on the efficiency of our method. The scatter-to-primary ratio estimation error on the Catphan(C)600 phantom without any annulus (20 cm in diameter) is at the level of 0.04, it rises to 0.07 and 0.1 on the phantom with an elliptical annulus (30 cm in the minor axis and 38 cm in the major axis) and with a circular annulus (38 cm in diameter). Conclusions: On the three phantom studies, good scatter correction performance of the proposed method has been demonstrated using both image comparisons and quantitative analysis. The theory and experiments demonstrate that a strong primary modulation that possesses a low transmission factor and a high modulation frequency is preferred for high scatter correction accuracy.
Keyword:
aluminium
biomedical materials
computerised tomography
copper
phantoms
X-ray imaging
X-ray scattering

期刊

Medical Physics 封面图
Medical Physics
IF:
3.2
论文数:
3.7W
被引数:
3.2W

机构

G
Georgia Institute of Technology
学者数:
1.8W
论文数: 1.4W
被引数: 5.9W
Q
qualcomm
学者数:
782
论文数: 646
被引数: 1
S
Stanford University
学者数:
9.6W
论文数: 8.2W
被引数: 17.0W
U
university system of georgia
学者数:
7.3W
论文数: 6.5W
被引数: 101
学者 查看更多机构
引用论文

引用论文

err分享
err收藏
Noise suppression in scatter correction for cone-beam CT
err2009-02-10
err149
errOAAI
errZhu, Lei; Wang, Jing; Xing, Lei
err分享
err收藏
err分享
err收藏
Hit ratios
err1976-01-01
err0
errOAAI
errS. J. Waters
err分享
err收藏
JUVENILE RHEUMATOID ARTHRITIS
err1996-08-01
err0
PREAI
errBarry P. Simmons; John T. Nutting; Richard A. Bernstein
err分享
err收藏
Scatter correction for cone-beam CT in radiation therapy
err2009-05-18
err194
errOAAI
errZhu, Lei; Xie, Yaoqin; Wang, Jing; Xing, Lei
err分享
err收藏
err分享
err收藏
学者 查看更多内容