返回
Sentinel lymph node fingerprinting
DOI:10.1088/1361-6560/ab177d.png)
摘要
En 中文
Background. When locating the sentinel lymph node (SLN), surgeons use state-of-the-art imaging devices, such as a 1D gamma probe or less widely spread a 2D gamma camera. These devices project the 3D subspace onto a 1D respectively 2D space, hence loosing accuracy and the depth of the SLN which is very important, especially in the head and neck area with many critical structures in close vicinity. Recent methods which use a multi-pinhole collimator and a single gamma detector image try to gain a depth estimation of the SLN. The low intensity of the sources together with the computational cost of the optimization process make the reconstruction in real-time, however, very challenging. Results. In this paper, we use an optimal design approach to improve the classical pinhole design, resulting in a non-symmetric distribution of the pinholes of the collimator. This new design shows a great improvement of the accuracy when reconstructing the position and depth of the radioactive tracer. Then, we introduce our Sentinel lymph node fingerprinting (SLNF) algorithm, inspired by MR-fingerprinting, for fast and accurate reconstruction of the tracer distribution in 3D space using a single gamma detector image. As a further advantage, the method requires no pre-processing, i.e. filtering of the detector image. The method is very stable in its performance even for low exposure times. In our ex vivo experiments, we successfully located multiple Technetium 99m (Tc-99m) sources with an exposure time of only one second and still, with a very small L-2-error. Conclusion. These promising results under short exposure time are very encouraging for SLN biopsy. Although, this device has not been tested on patients yet, we believe: that this approach will give the surgeon accurate 3D positions of the SLN and hence, can potentially reduce the trauma for the patient.
Keyword:
sentinel lymph node detection
fingerprinting
3D reconstruction
freehand SPECT
multi pinhole collimator
optimal design
optimal control
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
3.4
论文数:
1.4W
被引数:
3.5W
机构
引用论文
Alterations in complement-induced shape change and stimulus-specific superoxide anion generation by neonatal calf neutrophils
Inflammation
IF0

