Simultaneous misalignment correction for approximate circular cone-beam computed tomography
DOI:10.1088/0031-9155/53/22/001.png)
摘要
En 中文
Currently, CT scanning is often performed using flat detectors which are mounted on C-arm units or dedicated gantries as in radiation therapy or micro CT. For perspective cone-beam backprojection of the Feldkamp type (FDK) the geometry of an approximately circular scan trajectory has to be available for reconstruction. If the system or the scan geometry is afflicted with geometrical instabilities, referred to as misalignment, a non-perfect approximate circular scan is the case. Reconstructing a misaligned scan without knowledge of the true trajectory results in severe artefacts in the CT images. Unlike current methods which use a pre-scan calibration of the geometry for defined scan protocols and calibration phantoms, we propose a real-time iterative restoration of reconstruction geometry by means of entropy minimization. Entropy minimization is performed combining a simplex algorithm for multi- parameter optimization and iterative graphics card (GPU)-based FDK-reconstructions. Images reconstructed with the misaligned geometry were used as an input for the entropy minimization algorithm. A simplex algorithm changes the geometrical parameters of the source and detector with respect to the reduction of entropy. In order to reduce the size of the high-dimensional space required forminimization, the trajectory was described by only eight fix points. Avirtual trajectory is generated for each iteration using a least-mean-squares algorithm to calculate an approximately circular path including these points. Entropy was minimal for the ideal dataset, whereas strong misalignment resulted in a higher entropy value. For the datasets used in this study, the simplex algorithm required 64 - 200 iterations to achieve an entropy value equivalent to the ideal dataset, depending on the grade of misalignment using random initialization conditions. The use of the GPU reduced the time per iteration as compared to a quad core CPU-based backprojection by a factor of 10 resulting in a total of 15-20 ms per iteration, and thus providing an online geometry restoration after a total computation time of approximately 1-3 s, depending on the number of iterations. The proposed method provides accurate geometry restoration for approximately circular scans and eliminates the need for an elaborate off-line calibration for each scan. If a priori information about the trajectory is used to initialize the simplex algorithm, it is expected that the entropy minimization will converge significantly faster.
Keyword:
ARM MOUNTED XRII
C-ARM
GEOMETRICAL CALIBRATION
SIMPLEX-METHOD
RECONSTRUCTION
ALGORITHM
SYSTEM
PARAMETERS
DETECTOR
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
3.4
论文数:
1.4W
被引数:
3.1W
机构
引用论文
Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Image-based correction of gantry motion nonidealities
MEDICAL PHYSICS
IF3.2
The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator
MEDICAL PHYSICS
IF3.2
A Novel P-Selectin Glycoprotein Ligand-1 Monoclonal Antibody Recognizes an Epitope Within the Tyrosine Sulfate Motif of Human PSGL-1 and Blocks Recognition of Both P- and L-Selectin
Blood
IF0
Accurate technique for complete geometric calibration of cone-beam computed tomography systems锥束计算机断层扫描系统的完整几何校准的精确技术
MEDICAL PHYSICS
IF3.2