To What Extent Can the Current Amplitude Linearity of Rogowski Coils Be Verified?

delete2011-07-01
delete16
PRE
AI
B
Branislav Djokić *
J
J.D. Ramboz
D
D.E. Destefan
DOI:10.1109/TIM.2010.2099250delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
Calibrations of Rogowski coils at power frequencies are often performed at currents of only tens of amperes even though the coils are used at currents larger by two to three orders of magnitude, or even more. The assumption of an ideal linearity versus current amplitude is true for ideal coils. This paper addresses the question on practical significance as to what extent the current amplitude linearity of real Rogowski coils can be verified at power frequencies. For this task, several different types of Rogowski coils were evaluated using a digital sampling system, with the best calibration uncertainties (k = 2) of 50 mu A/A for magnitude and 50 mu rad for phase and the estimated linearity better by a factor of two or more.
Keywords:
AC current measurement
calibration
digital sampling
linearity
Rogowski coil
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

IEEE Transactions on Instrumentation and Measurement cover
IEEE Transactions on Instrumentation and Measurement
IF:
5.9
Papers:
1.9W
Citations:
5.8W

Organization

N
National Research Council Canada
Scholars:
7.9K
Papers: 7.9K
Citations: 6.8K
Cited Papers

Cited Papers

Citing Papers

Citing Papers