Volumetric Real-Time Imaging Using a CMUT Ring Array
DOI:10.1109/TUFFC.2012.2310.png)
摘要
En 中文
A ring array provides a very suitable geometry for forward-looking volumetric intracardiac and intravascular ultrasound imaging. We fabricated an annular 64-element capacitive micromachined ultrasonic transducer (CMUT) array featuring a 10-MHz operating frequency and a 1.27-mm outer radius. A custom software suite was developed to run on a PC-based imaging system for real-time imaging using this device. This paper presents simulated and experimental imaging results for the described CMUT ring array. Three different imaging methods-flash, classic phased array (CPA), and synthetic phased array (SPA)-were used in the study. For SPA imaging, two techniques to improve the image quality-Hadamard coding and aperture weighting-were also applied. The results show that SPA with Hadamard coding and aperture weighting is a good option for ring-array imaging. Compared with CPA, it achieves better image resolution and comparable signal-to-noise ratio at a much faster image acquisition rate. Using this method, a fast frame rate of up to 463 volumes per second is achievable if limited only by the ultrasound time of flight; with the described system we reconstructed three cross-sectional images in real-time at 10 frames per second, which was limited by the computation time in synthetic beamforming.
Keyword:
MICROMACHINED ULTRASONIC TRANSDUCERS
HIGH-FRAME RATE
FABRICATION
COHERENT
期刊
IF:
3.7
论文数:
6.7K
被引数:
1.4W
机构
引用论文
População negra: o que não é ensinado sobre semiologia dermatológica na pele preta - Um relato de experiência黑人人口:关于黑色皮肤皮肤病学症状学未被教授的内容——一份经验报告
THE UNIFYING ROLE OF THE COARRAY IN APERTURE SYNTHESIS FOR COHERENT AND INCOHERENT IMAGING
PROCEEDINGS OF THE IEEE
IF25.9