返回
Waveguide permittivity measurement using variable-length samples and an uncalibrated reflectometer
DOI:10.1109/19.481351.png)
摘要
En 中文
A method is presented to measure the complex permittivity of samples inserted into waveguides. The method is based on reflection coefficient measurements for four different sample lengths, which are used to determine the complex propagation constant of the guide's fundamental mode. For simple waveguide geometries, the complex permittivity can be calculated, assuming a value of 1 for the relative permeability. Unlike classical waveguide techniques, this method does not require calibration of the reflectometer, and it can be used with samples that are narrower than the waveguide cross section's main dimension, By using narrower samples, the upper frequency limit of the method can be increased. Measurements in an S-band rectangular waveguide using the proposed method are presented and compared with results obtained with a cavity perturbation method, Very good agreement has been found over the frequency range prescribed for the method, Results are shown for completely filled and partially filled waveguides.
Keyword:
PERMEABILITY
AI总结
对已上传原文的论文进行重点信息的提取,主要内容包括:简要概述、研究摘要、背景介绍、关键亮点、图文解析、展望与总结。
期刊
IF:
5.9
论文数:
2.0W
被引数:
5.8W
机构
暂无机构信息
引用论文
AUTOMATIC MEASUREMENT OF COMPLEX DIELECTRIC-CONSTANT AND PERMEABILITY AT MICROWAVE-FREQUENCIES
PROCEEDINGS OF THE IEEE
IF25.9