A capillary-driven microfluidic device for performing spatial multiplex PCR
DOI:10.1007/s10544-025-00745-2.png)
Abstract
En 中文
Multiplex polymerase chain reaction (PCR) tests multiple biomarkers or pathogens that cause overlapping symptoms, making it an essential tool in syndromic testing. To achieve a multiplex PCR on chip, a design based on capillary-driven fluidic actuation is proposed. Our silicon chip features 22 reaction chambers and allows primers and probes to be pre-spotted in the reaction chambers prior to use. The design facilitates rapid sample loading through a common inlet channel, delivering reagents to all reaction chambers in less than 10 s. A custom clamping mechanism combined with a double depth cavity design ensures proper sealing during temperature cycling without the need for extra reagents like oil. Temperature cycling and fluorescence imaging were performed using custom-made hardware. As a proof of concept, two single nucleotide polymorphisms (SNPs), CyP2C19*2 and PCSK9 were detected. These results demonstrate the feasibility of on-chip multiplex PCR, compatible with different assays in parallel and requiring only a single pipetting step for reagent loading, without active fluidic actuation like pumping.
Keywords:
Capillary flow
Multiplex PCR
Silicon chip
Lab-on-a-chip
Fluorescence detection