arrow
Return

A Self-Charging Supercapacitor for a Patch-Type Glucose Sensor

delete2022-01-13
delete32
PRE
AI
H
Hye-Jun Kil
S
Seung‐Rok Kim
J
Jin‐Woo Park *
DOI:10.1021/acsami.1c21394delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
Wearable electronic medical devices measuring continuous biological signals for early disease diagnosis should be small and lightweight for consecutive usability. As a result, there has been an increasing need for new energy supply systems that provide continuous power without any interruption to the operation of the medical devices associated with the use of conventional batteries. In this work, we developed a patch-type self-charging supercapacitor that can measure biological signals with a continuous energy supply without batteries. The glucose oxidase coated on the surface of the microneedle-type glucose sensor encounters glucose in the interstitial fluids of the human body. Electrons created by glucose oxidation operate the self-powered system in which charging begins with the generation of potential differences in supercapacitor electrodes. In an 11 mM glucose solution, the self-powered solidstate supercapacitors (SPSCs) showed a power density of 0.62 mW/cm(2), which resulted in self-charging of the supercapacitor. The power density produced by each SPSC with a drop of 11 mM glucose solution was higher than that produced by glucose-based biofuel cells. Consequently, the all-in-one self-powered glucose sensor, with the aid of an Arduino Uno board and appropriate programming, effectively distinguished normal, prediabetic, and diabetic levels from 0.5 mL of solutions absorbed in a laboratory skin model.
Keywords:
supercapacitor
self-charging
microneedle
enzyme
glucose sensor
self-powered system

Journal

ACS Applied Materials and Interfaces cover
ACS Applied Materials and Interfaces
IF:
8.2
Papers:
6.1W
Citations:
38.7W

Organization

Y
Yonsei University
Scholars:
4.8W
Papers: 4.6W
Citations: 5.2W