Return
A Temperature-Differential Affinity Biosensor: Model and D-Optimal Performance Limits
DOI:10.1109/JSEN.2011.2106156.png)
Abstract
En 中文
Selectivity is one of the most challenging issues in biosensor design. Several methods have been proposed in the past to overcome nonspecific interference. In particular, it has been shown that temperature curves can be used to simultaneously measure two similar analytes. Here, the performance of such thermal-based systems is analyzed, using least squares estimation and existing models for affinity-based sensing, and it is shown that a D-optimal difference for the sensor temperatures exists. Analysis at this optimized condition yields bounds on the sensitivity and selectivity of this class of sensors. For the first time, thermal discrimination is employed for an affinity-based sensor: an artificial receptor-based system for the detection of the neurotransmitter acetylcholine and its metabolite choline. The system performance demonstrates the practicality of the theoretical results presented.
Keywords:
Biosensor
interference elimination
simultaneous measurement
AI Summary
Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.
Journal
IF:
4.5
Papers:
2.2W
Citations:
7.3W
Organization
Cited Papers
Weighted least squares approach to calculating limits of detection and quantification by modeling variability as a function of concentration
ANALYTICAL CHEMISTRY
IF6.7