arrow
Return

Anatomy-based three-dimensional dose optimization in brachytherapy using multiobjective genetic algorithms

delete1999-09-07
delete65
PRE
AI
M
Michael Lahanas
D
Dimos Baltas
N
Nikolaos Zamboglou
DOI:10.1118/1.598697delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
In conventional dose optimization algorithms, in brachytherapy, multiple objectives are expressed in terms of an aggregating function which combines individual objective values into a single utility value, making the problem single objective, prior to optimization. A multiobjective genetic algorithm (MOGA) was developed for dose optimization based on an a posteriori approach, leaving the decision-making process to a planner and offering a representative trade-off surface of the various objectives. The MOGA provides a flexible search engine which provides the maximum of information for a decision maker. Tests performed with various treatment plans in brachytherapy have shown that MOGA gives solutions which are superior to those of traditional dose optimization algorithms. Objectives were proposed in terms of the COW distribution and differential volume histograms, taking into account patient anatomy in the optimization process. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00309-0].
Keywords:
optimization
brachytherapy
COIN
dose-volume histograms
multiobjective genetic algorithms
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

Medical Physics cover
Medical Physics
IF:
3.2
Papers:
3.7W
Citations:
3.2W

Organization

No organization information available
Cited Papers

Cited Papers

errShare
errSave
A new genetic algorithm technique in optimization of permanent 125I prostate implants
err1998-12-03
err71
PREAI
errYang, GZ; Reinstein, LE; Pai, S; Xu, ZG; Carroll, DL
errShare
errSave
A conformal index (COIN) to evaluate implant quality and dose specification in brachytherapy
err1998-01-01
err308
PREAI
errBaltas, D; Kolotas, C; Geramani, K; Mould, RF; Ioannidis, G; Kekchidi, M; Zamboglou, N
errShare
errSave
6‐Methoxy‐1,2,3,4‐tetrahydro‐β‐carboline–α serotonin elevator1
err2006-10-04
err0
PREAI
errW. M. McIsaac; Dorothy Taylor; K. E. Walker; B. T. Ho
errShare
errSave
errShare
errSave
researcher View more