Return
Autonomous microfluidic system for phosphate detection
DOI:10.1016/j.talanta.2006.06.011.png)
Abstract
En 中文
Miniaturization of analytical devices through the advent of microfluidics and micro total analysis systems is an important step forward for applications such as medical diagnostics and environmental monitoring. The development of field-deployable instruments requires that the entire system, including all necessary peripheral components, be miniaturized and packaged in a portable device. A sensor for long-term monitoring of phosphate levels has been developed that incorporates sampling, reagent and waste storage, detection, and wireless communication into a complete, miniaturized system. The device employs a low-power detection and communication system, so the entire instrument can operate autonomously for 7 days on a single rechargeable, 12 V battery. In addition, integration of a wireless communication device allows the instrument to be controlled and results to be downloaded remotely. This autonomous system has a limit of detection of 0.3 mg/L and a linear dynamic range between 0 and 20 mg/L. d (c) 2006 Elsevier B.V. All rights reserved.
Keywords:
wireless
sensor
network
environment
phosphate
AI Summary
Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.
Journal
IF:
6.1
Papers:
2.6W
Citations:
6.3W
Organization
No organization information available
Cited Papers
No cited papers available