Biomechanical Evaluation of Anterior Plate Fixation With Cage for Basilar Invagination With Atlantoaxial Dislocation: A Cadaveric Study

delete2025-12-01
delete0
delete
OA
AI
J
Jianying Zheng
Z
Zhiping Huang
K
K. F. ng Li
X
X. Zhang
J
Jian Xiong
Y
Yu Wang
J
Jiahao Xie
P
Panjie Xu
Z
Zhongmin Zhang
W
Wei Ji *
DOI:10.14245/ns.2550694.347delete
deleteOriginal
deleteShare
deleteSave
View PDF
Abstract

Abstract

En 中文
Objective: To evaluate the biomechanical characteristics of 2 anterior fixation techniques occipitocervical fixation (POCF) for basilar invagination with atlantoaxial dislocation (BIAAD), under varying atlantoaxial lateral mass cage heights (4-10 mm). Methods: Seven fresh cadaveric specimens (occiput to C3, Oc-C3) were tested in the following conditions: (1) intact state; (2) BI-AAD state; (3) BI-AAD+CPF; (4) BI-AAD+TARP fixation; (5) BI-AAD+POCF. A pure 1.5 Nm moment loads to specimens in flexion/extension, lateral bending and axial rotation. Range of motion (ROM) and neutral zone (NZ) values at Oc-C2 were calculated and compared. Results: ROM of the C1-2 segment under the intact and BI-AAD states were as follows: 9.3 degrees +/- 4.6 degrees versus 21.3 degrees +/- 8.3 degrees in flexion, 4.6 degrees +/- 1.9 degrees versus 9.3 degrees +/- 3.8 degrees in extension, 3.6 degrees +/- 2.2 degrees versus 12.0 degrees +/- 6.5 degrees in lateral bending, and 68.9 degrees +/- 14.4 degrees versus 76.6 degrees +/- 6.6 degrees in axial rotation, respectively. Compared with BI-AAD states, all internal fixation techniques significantly reduced the ROM of the Oc-C2 segment. TARP fixation exhibited larger ROM in flexion-extension. While in lateral bending and axial rotation, the ROM values for the anterior plate constructs were smaller than that of POCF, with a statistically significant difference observed between CPF and POCF. Cage height variations showed no significant impact on overall biomechanical stability. Conclusion: Anterior plate fixation techniques demonstrated superior resistance to lateral bending and rotational forces compared to posterior approaches, with clival plate fixation exhibiting optimal biomechanical stability for BI-AAD. Variations in cage height exhibited negligible impact on stability when internal fixation achieved adequate rigidity.
Keywords:
Basilar invagination
Atlantoaxial dislocation
Internal fixation
Biomechanics
Fusion cage
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

N
Neurospine
IF:
3.6
Papers:
920
Citations:
2.0K

Organization

S
southern medical university - china
Scholars:
4.5W
Papers: 2.4W
Citations: 50
Cited Papers

Cited Papers

Citing Papers

Citing Papers