Calibration Method for Reducing Ring Artifacts in Energy Bin Images for Photon-Counting CT

delete2025-08-28
delete0
PRE
AI
D
Donghyeon Lee
X
Xiaohui Zhan
W
Wen-Hsin Tai
S
Shalini Subramanian
W
Wojciech Zbijewski
K
Katsuyuki Taguchi
DOI:10.1109/TMI.2025.3603880delete
deleteOriginal
deleteShare
deleteSave
Abstract

Abstract

En 中文
Pixel-to-pixel variation in outputs of photon-counting detectors (PCDs) is a significant challenge in computed tomography (CT) imaging, leading to ring artifacts. This study presents a novel calibration framework to address this issue. Our approach involves a streamlined data acquisition strategy using CT scans of calibration phantoms, enabling comprehensive phantom thickness data acquisition across all pixels, coupled with three dedicated calibration steps. This process generates two correction coefficient tables: one for pixel-to-pixel variation correction and the other for both pixel-to-pixel variation and additional count-rate non-linearity correction using water as the calibration material. For experimental validation, we utilized a prototype PCD-based CT system and quantitatively evaluated the performance with high-frequency (ring artifact) and low-frequency (cupping/doming artifact) radial mean standard deviations (RMSDs) on both total count data and individual energy bin count data. Compared to a conventional calibration method, the proposed approach achieved substantial reductions in RMSDs, with improvements of 95.7% for total counts and 93.6% for energy bin counts in high-frequency RMSDs, and 97.2% for total counts and 55.7% for energy bin counts in low-frequency RMSDs. We anticipate that this calibration method will serve as an effective solution for addressing pixel-to-pixel variation in PCDs.
Keywords:
Pixel-to-pixel variation
ring artifact
calibration
photon-counting detector
computed tomography

Journal

IEEE Transactions on Medical Imaging cover
IEEE Transactions on Medical Imaging
IF:
9.8
Papers:
6.4K
Citations:
3.7W

Organization

J
johns hopkins university
Scholars:
1.1W
Papers: 4.1K
Citations: 1
C
canon medical research usa inc
Scholars:
1
Papers: 1
Citations: 0
Cited Papers

Cited Papers

errShare
errSave
An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors
err2010-07-12
err260
errOAAI
errTaguchi, Katsuyuki; Frey, Eric C.; Wang, Xiaolan; Iwanczyk, Jan S.; Barber, William C.
errShare
errSave
Log conversion with fewer counts for photon counting CT: decreasing both bias and variance simultaneously
err2025-04-08
err0
PREAI
errWei,Jia; Lee,Donghyeon; Stierstorfer,Karl; Fung,George S. K.; Polster,Christoph; Subramanian,Shalini; Taguchi,Katsuyuki
errShare
errSave
A Ring Artifact Correction Method: Validation by Micro-CT Imaging with Flat-Panel Detectors and a 2D Photon-Counting Detector
errSENSORS
IF3.5
err2017-01-30
err37
errOAAI
errEldib, Mohamed Elsayed; Hegazy, Mohamed; Mun, Yang Ji; Cho, Myung Hye; Cho, Min Hyoung; Lee, Soo Yeol
errShare
errSave
errShare
errSave
Clinical Applications of Photon-counting CT: A Review of Pioneer Studies and a Glimpse into the Future
errRADIOLOGY
IF15.2
err2023-10-01
err27
errOAAI
errDouek, Philippe C.; Boccalini, Sara; Oei, Edwin H. G.; Cormode, David P.; Pourmorteza, Amir; Boussel, Loic; Si-Mohamed, Salim A.; Budde, Ricardo P. J.
errShare
errSave
errShare
errSave
researcher View more