arrow
Return

Complex Permittivity Measurement Utilizing Multiple Modes of a Rectangular Cavity

delete2023-01-01
delete9
PRE
AI
Q
Qin Shi
Q
Qing‐Xin Chu *
M
Meizhen Xiao
陈付昌 (Fu‐Chang Chen)
X
Xue-Quan Huang
X
Xiao He
DOI:10.1109/TIM.2022.3225926delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
A novel method for measuring the dielectric constant and loss tangent of dielectric substrates is presented in this article. Four resonant modes of the rectangular resonant cavity, TE101, TE105, TE109, and TE1015, are utilized to realize the multiband measurement from 0.9 to 10 GHz. The introduction of additional slots cuts down the quality factor of some miscellaneous modes, making the working modes more obvious and convenient for extraction. Based on the traditional perturbation formulas, considering the sinusoidal field distribution in the sample, modified formulas are deduced to obtain more accurate results. An experimental resonant cavity is fabricated and measured to demonstrate the effectivity of the proposed method.
Keywords:
Resonant frequency
Magnetic field measurement
Permittivity measurement
Perturbation methods
Frequency measurement
Dielectric loss measurement
Loss measurement
Complex permittivity
multiband measurement
perturbation
rectangular cavity resonator

Journal

IEEE Transactions on Instrumentation and Measurement cover
IEEE Transactions on Instrumentation and Measurement
IF:
5.9
Papers:
1.9W
Citations:
5.8W

Organization

S
south china university of technology
Scholars:
6.7W
Papers: 5.1W
Citations: 85