arrow
Return

Correcting inter-scan motion artifacts in quantitative R1 mapping at 7T

delete2022-03-21
delete3
delete
OA
AI
Y
Yaël Balbastre
A
Ali Aghaeifar
N
Nadège Corbin
M
Mikael Brudfors
J
John Ashburner
M
Martina F. Callaghan *
DOI:10.1002/mrm.29216delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
Purpose: Inter-scan motion is a substantial source of error in R1 estimation methods based on multiple volumes, for example, variable flip angle (VFA), and can be expected to increase at 7T where B1 fields are more inhomogeneous. The established correction scheme does not translate to 7T since it requires a body coil reference. Here we introduce two alternatives that outperform the established method. Since they compute relative sensitivities they do not require body coil images. Theory: The proposed methods use coil-combined magnitude images to obtain the relative coil sensitivities. The first method efficiently computes the relative sensitivities via a simple ratio; the second by fitting a more sophisticated generative model. Methods: R1 maps were computed using the VFA approach. Multiple datasets were acquired at 3T and 7T, with and without motion between the acquisition of the VFA volumes. R1 maps were constructed without correction, with the proposed corrections, and (at 3T) with the previously established correction scheme. The effect of the greater inhomogeneity in the transmit field at 7T was also explored by acquiring B+ 1 maps at each position. Results: At 3T, the proposed methods outperform the baseline method. Inter- scan motion artifacts were also reduced at 7T. However, at 7T reproducibility only converged on that of the no motion condition if position- specific transmit field effects were also incorporated. Conclusion: The proposedmethods simplify inter-scanmotion correction of R1 maps and are applicable at both 3T and 7T, where a body coil is typically not available. The open-source code for all methods is made publicly available.
Keywords:
7T
generative modeling
inter-scan motion
sensitivity
qMRI
R-1

Journal

Magnetic Resonance in Medicine cover
Magnetic Resonance in Medicine
IF:
3
Papers:
1.2W
Citations:
3.1W

Organization

U
university of london
Scholars:
21.5W
Papers: 19.7W
Citations: 305