Return
Design and performance of a miniaturized cesium free-induction-decay optically pumped magnetometer with an integrated polarization–heating probe
R
H
X
T
C
DOI:10.1016/j.sna.2026.118306.png)
Abstract
En 中文
• An integrated polarization-heating probe is developed for Cs FID magnetometry. • A single helical coil provides vapor-cell heating and polarization. • Temporal decoupling suppresses coil-induced perturbations during FID readout. • The 18.5 × 17.5 × 36.5 mm probe consumes about 0.7 W. • The sensor achieves 15 pT/√Hz sensitivity with a maximum residual of 8.980 nT after correction.
Keywords:
Polarization–heating integration
Free-induction decay (FID)
Cesium optically pumped magnetometer
Low power consumption
Miniaturization
Journal
IF:
4.9
Papers:
1.5W
Citations:
3.3W