Eddy Current Probe Identification and Analysis
DOI:10.1109/TIM.2017.2684858.png)
Abstract
En 中文
This paper presents a new approach to the classical method for failure detection with eddy currents. In contrast to traditional methods of measuring reflected impedance on the probe circuit, this paper uses concepts of identification theory to find a transfer function that characterizes the dynamics of the measuring system by eddy currents. The transfer function represents the admittance of the input equivalent circuit. The estimated parameters of the transfer function were used to compute the inductive time constant of equivalent circuit of the metal specimen. The results of the time constant variation of the equivalent circuit of the sample were compared with the equivalent input impedance variation. A preliminary analysis is done in a simple case study, which has shown that the inductive time constant is more sensitive than the reflected impedance on the probe circuit, especially at low frequencies. Moreover, the estimated time constant is independent of the excitation frequency and mutual inductance.
Keywords:
Eddy currents
nondestructive testing
sensitivity
system identification