Evaluating a CBCT Correction Algorithm for Adaptive Radiotherapy
DOI:10.1177/15330338261428647.png)
Abstract
En 中文
Introduction
To increase efficiency of adaptive radiotherapy (ART), we tested a cone beam computed tomography (CBCT) correction algorithm to evaluate the feasibility of utilizing daily CBCTs for treatment planning.
Methods
A lung phantom was scanned with a CT and CBCT on two different linacs. The CBCTs were processed through a correction algorithm in the treatment planning system (TPS). The algorithm reduces artifacts and adjusts image intensity to more closely match the planning CT, to generate corrected CBCTs. Voxels outside the CBCT field of view (FOV) are replaced with voxels from the planning CT. A treatment plan was first generated on the CT, then recalculated on the corrected CBCTs. The same workflow was followed for seven previously adapted head and neck and seven sarcoma patients. Each patient's adaptive plan was recalculated on the corrected CBCTs. Dose differences were analyzed for these plans using a 3%/2 mm gamma analysis.
Results
Both Ethos and TrueBeam CBCT plans on the phantom had high matching dose per voxel according to gamma analysis. After corrections of some registration errors, all 14 plans achieved gamma passing rate above 95% (3%/2 mm).
Conclusions
The CBCT correction algorithm demonstrates potential to reduce the need for re-simulation and enable faster offline adaptive planning without sacrificing dose calculation accuracy.
Keywords:
CBCT correction algorithm
Adaptive radiotherapy
Dose calculation accuracy
Image intensity adjustment
Gamma analysis
Journal
T
IF:
2.8
Papers:
93
Citations:
0
Organization
Cited Papers
Daily dose evaluation based on corrected CBCTs for breast cancer patients: accuracy of dose and complication risk assessment
RADIATION ONCOLOGY
IF3.2
An optimized framework for cone-beam computed tomography-based online evaluation for proton therapy
MEDICAL PHYSICS
IF3.2