arrow
Return

Graphene-based microelectrodes with bidirectional functionality for next-generation retinal electronic interfaces

delete2024-01-01
delete2
delete
OA
AI
F
Fikret Taygun Duvan
M
Marina Cunquero
E
Eduard Masvidal‐Codina
S
Steven T. Walston
M
María Marsal
J
Jose Manuel de la Cruz
D
Damià Viana
D
Diep Nguyen
J
Julie Dégardin
X
Xavi Illa
J
J Zhang
M
María del Pilar Bernicola
J
José Gabriel Macias-Montero
G
Gustavo Castro‐Olvera
E
Elena del Corro
S
Socrates Dokos
M
M. Chmeissani
P
Pablo Loza‐Álvarez
S
Serge Picaud
J
José A. Garrido *
DOI:10.1039/d4nh00282bdelete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
Neuroelectronic prostheses are being developed for restoring vision at the retinal level in patients who have lost their sight due to photoreceptor loss. The core component of these devices is the electrode array, which enables interfacing with retinal neurons. Generating the perception of meaningful images requires high-density microelectrode arrays (MEAs) capable of precisely activating targeted retinal neurons. Achieving this precision necessitates the downscaling of electrodes to micrometer dimensions. However, miniaturization increases electrode impedance, which poses challenges by limiting the amount of current that can be delivered, thereby impairing the electrode's capability for effective neural modulation. Additionally, it elevates noise levels, reducing the signal quality of the recorded neural activity. This report focuses on evaluating reduced graphene oxide (rGO) based devices for interfacing with the retina, showcasing their potential in vision restoration. Our findings reveal low impedance and high charge injection limit for microscale rGO electrodes, confirming their suitability for developing next-generation high-density retinal devices. We successfully demonstrated bidirectional interfacing with cell cultures and explanted retinal tissue, enabling the identification and modulation of multiple cells' activity. Additionally, calcium imaging allowed real-time monitoring of retinal cell dynamics, demonstrating a significant reduction in activated areas with small-sized electrodes. Overall, this study lays the groundwork for developing advanced rGO-based MEAs for high-acuity visual prostheses. This article presents the development and testing of reduced graphene oxide (rGO) microelectrode arrays for interfacing with retinal neurons, showcasing their potential for high-density visual prosthetics.
Keywords:
ELECTRICAL-STIMULATION
TEMPORAL PROPERTIES
VISUAL-PERCEPTION
GANGLION-CELLS
TRIAL
MODEL

Journal

Nanoscale Horizons cover
Nanoscale Horizons
IF:
6.6
Papers:
1.3K
Citations:
6.3K

Organization

B
barcelona institute of science & technology
Scholars:
1.2W
Papers: 9.7K
Citations: 36
C
consejo superior de investigaciones cientificas (csic)
Scholars:
8.8W
Papers: 8.5W
Citations: 125
U
universitat politecnica de catalunya
Scholars:
1.9W
Papers: 1.6W
Citations: 17
A
Autonomous University of Barcelona
Scholars:
3.7W
Papers: 2.6W
Citations: 47
researcher View more organizations