Imaging Performance for Two Row-Column Arrays

delete2019-07-01
delete34
delete
OA
AI
H
Hamed Bouzari
M
Mathias Engholm
S
Svetoslav Ivanov Nikolov
M
Matthias Bo Stuart
E
Erik Vilain Thomsen
J
Jørgen Arendt Jensen *
DOI:10.1109/TUFFC.2019.2914348delete
deleteOriginal
deleteShare
deleteSave
View PDF
Abstract

Abstract

En 中文
This study evaluates the volumetric imaging performance of two prototyped 62 + 62 row-column-addressed (RCA) 2-D array transducer probes using three synthetic aperture imaging (SAI) emission sequences and two different beamformers. The probes are fabricated using capacitive micromachined ultrasonic transducer (CMUT) and piezoelectric transducer (PZT) technology. Both have integrated apodization to reduce ghost echoes and are designed with similar acoustical features, i.e., 3-MHz center frequency, lambda/2 pitch, and 24.8 x 24.8 mm(2) active footprint. Raw RF data are obtained using an experimental research ultrasound scanner, SARUS. The SAI sequences are designed for imaging down to 14 cm at a volume rate of 88 Hz. Two beamforming methods, spatial matched filtering and row-column adapted delay-and-sum, are used for beamforming the RF data. The imaging quality is investigated through simulations and phantom measurements. Both probes on average have similar lateral full-width at half-maximum (FWHM) values, but the PZT probe has 20% smaller cystic resolution values and 70% larger contrast-to-noise ratio (CNR) compared to the capacitive micromachined ultrasonic transducer (CMUT) probe. The CMUT probe can penetrate down to 15 cm, and the PZT probe down to 30 cm. The CMUT probe has 17% smaller axial FWHM values. The matched filter focusing shows an improved B-mode image for measurements on a cyst phantom with an improved speckle pattern and better visualization of deeper lying cysts. The results of this study demonstrate the potentials of RCA 2-D arrays against fully addressed 2-D arrays, which are low channel count (e.g., 124 instead of 3844), low acoustic intensity mechanical index (MI <= 0.88 and spatial-peak-temporal-average intensity I-spta <= 5.5 mW/cm(2)), and high penetration depth (down to 30 cm), which makes 3-D imaging at high volume rates possible with equipment in the price range of conventional 2-D imaging.
Keywords:
3-D imaging
beamforming
row-column arrays
volumetric ultrasound imaging
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control cover
IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
IF:
3.7
Papers:
6.7K
Citations:
1.4W

Organization

T
technical university of denmark
Scholars:
2.6W
Papers: 2.8W
Citations: 37