Improving apparent diffusion coefficient accuracy on a compact 3T MRI scanner using gradient nonlinearity correction

delete2018-09-26
delete13
delete
OA
AI
A
Ashley Tao
Y
Yunhong Shu *
E
Ek T. Tan
J
Joshua D. Trzasko
S
Shengzhen Tao
R
Robert D. Reid
P
Paul T. Weavers
J
John Huston
M
Matt A. Bernstein
DOI:10.1002/jmri.26201delete
deleteOriginal
deleteShare
deleteSave
View PDF
Abstract

Abstract

En 中文
Background Gradient nonlinearity (GNL) leads to biased apparent diffusion coefficients (ADCs) in diffusion-weighted imaging. A gradient nonlinearity correction (GNLC) method has been developed for whole body systems, but is yet to be tested for the new compact 3T (C3T) scanner, which exhibits more complex GNL due to its asymmetrical design. Purpose To assess the improvement of ADC quantification with GNLC for the C3T scanner. Study Type Phantom measurements and retrospective analysis of patient data. Phantom/Subjects A diffusion quality control phantom with vials containing 0-30% polyvinylpyrrolidone in water was used. For in vivo data, 12 patient exams were analyzed (median age, 33). Field Strength/Sequence Imaging was performed on the C3T and two commercial 3T scanners. A clinical DWI (repetition time [TR] = 10,000 msec, echo time [TE] = minimum, b = 1000 s/mm(2)) sequence was used for phantom imaging and 10 patient cases and a clinical DTI (TR = 6000-10,000 msec, TE = minimum, b = 1000 s/mm(2)) sequence was used for two patient cases. Assessment The 0% vial was measured along three orthogonal axes, and at two different temperatures. The ADC for each concentration was compared between the C3T and two whole-body scanners. Cerebrospinal fluid and white matter ADCs were quantified for each patient and compared to values in literature. Statistical Tests Paired t-test and two-way analysis of variance (ANOVA). Results For all PVP concentrations, the corrected ADC was within 2.5% of the reference ADC. On average, the ADC of cerebrospinal fluid and white matter post-GNLC were within 1% and 6%, respectively, of values reported in the literature and were significantly different from the uncorrected data (P < 0.05). Data Conclusion This study demonstrated that GNL effects were more severe for the C3T due to the asymmetric gradient design, but our implementation of a GNLC compensated for these effects, resulting in ADC values that are in good agreement with values from the literature.
Keywords:
diffusion weighted imaging
gradient non-linearity
apparent diffusion coefficient
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

Journal of Magnetic Resonance Imaging cover
Journal of Magnetic Resonance Imaging
IF:
3.5
Papers:
9.9K
Citations:
2.0W

Organization

G
General Electric
Scholars:
4.4K
Papers: 3.4K
Citations: 2
M
mayo clinic
Scholars:
8.3W
Papers: 6.5W
Citations: 85