Intraoperative microseizure detection using a high-density micro-electrocorticography electrode array

delete2022-05-27
delete16
delete
OA
AI
J
James Sun
K
Katrina Barth
S
Shaoyu Qiao
C
Chia‐Han Chiang
C
Charles Wang
S
Shervin Rahimpour
M
Michael Trumpis
S
Suseendrakumar Duraivel
A
Agrita Dubey
K
Katie Wingel
I
Iakov Rachinskiy
V
Voinas, Alex S.
B
Breonna Ferrentino
D
Derek G. Southwell
M
Michael M. Haglund
A
Allan H. Friedman
S
Shivanand P. Lad
W
Werner Doyle
F
Florian Solzbacher
G
Gregory B. Cogan
S
Saurabh R. Sinha
S
Sasha Devore
O
Orrin Devinsky
D
Daniel Friedman
B
Bijan Pesaran *
J
Jonathan Viventi *
DOI:10.1093/braincomms/fcac122delete
deleteOriginal
deleteShare
deleteSave
View PDF
Abstract

Abstract

En 中文
One-third of epilepsy patients suffer from medication-resistant seizures. While surgery to remove epileptogenic tissue helps some patients, 30-70% of patients continue to experience seizures following resection. Surgical outcomes may be improved with more accurate localization of epileptogenic tissue. We have previously developed novel thin-film, subdural electrode arrays with hundreds of microelectrodes over a 100-1000 mm(2) area to enable high-resolution mapping of neural activity. Here, we used these high-density arrays to study microscale properties of human epileptiform activity. We performed intraoperative micro-electrocorticographic recordings in nine patients with epilepsy. In addition, we recorded from four patients with movement disorders undergoing deep brain stimulator implantation as non-epileptic controls. A board-certified epileptologist identified microseizures, which resembled electrographic seizures normally observed with clinical macroelectrodes. Recordings in epileptic patients had a significantly higher microseizure rate (2.01 events/min) than recordings in non-epileptic subjects (0.01 events/min; permutation test, P = 0.0068). Using spatial averaging to simulate recordings from larger electrode contacts, we found that the number of detected microseizures decreased rapidly with increasing contact diameter and decreasing contact density. In cases in which microseizures were spatially distributed across multiple channels, the approximate onset region was identified. Our results suggest that micro-electrocorticographic electrode arrays with a high density of contacts and large coverage are essential for capturing microseizures in epilepsy patients and may be beneficial for localizing epileptogenic tissue to plan surgery or target brain stimulation. Sun, Barth et al. report that microseizures occur at higher rates in patients with epilepsy compared with patients without epilepsy in an intraoperative setting. They find that arrays of densely spaced microcontacts are essential to detecting microseizures, which may be used to improve the localization of epileptogenic tissue.
Keywords:
epilepsy
microseizure
ECoG
intraoperative
microelectrode
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

B
Brain Communications
IF:
4.5
Papers:
2.6K
Citations:
5.8K

Organization

D
Duke University
Scholars:
6.3W
Papers: 5.7W
Citations: 6.5W
N
New York University
Scholars:
4.4W
Papers: 3.9W
Citations: 5.8W
U
Utah System of Higher Education
Scholars:
4.6W
Papers: 4.0W
Citations: 161
researcher View more organizations