arrow
Return

Minimizing electric fields and increasing peripheral nerve stimulation thresholds using a body gradient array coil

delete2024-04-16
delete1
delete
OA
AI
R
Reza Babaloo
E
Ergin Atalar *
DOI:10.1002/mrm.30109delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
PurposeTo demonstrate the performance of gradient array coils in minimizing switched-gradient-induced electric fields (E-fields) and improving peripheral nerve stimulation (PNS) thresholds while generating gradient fields with adjustable linearity across customizable regions of linearity (ROLs).MethodsA body gradient array coil is used to reduce the induced E-fields on the surface of a body model by modulating applied currents. This is achieved by performing an optimization problem with the peak E-field as the objective function and current amplitudes as unknown variables. Coil dimensions and winding patterns are fixed throughout the optimization, whereas other engineering metrics remain adjustable. Various scenarios are explored by manipulating adjustable parameters.ResultsThe array design consistently yields lower E-fields and higher PNS thresholds across all scenarios compared with a conventional coil. When the gradient array coil generates target gradient fields within a 44-cm-diameter spherical ROL, the maximum E-field is reduced by 10%, 18%, and 61% for the X, Y, and Z gradients, respectively. Transitioning to a smaller ROL (24 cm) and relaxing the gradient linearity error results in further E-field reductions. In oblique gradients, the array coil demonstrates the most substantial reduction of 40% in the Z-Y direction. Among the investigated scenarios, the most significant increase of 4.3-fold is observed in the PNS thresholds.ConclusionOur study demonstrated that gradient array coils offer a promising pathway toward achieving high-performance gradient coils regarding gradient strength, slew rate, and PNS thresholds, especially in scenarios in which linear magnetic fields are required within specific target regions.
Keywords:
adjustable region of linearity
electromagnetic field simulation
flexible gradient linearity
gradient array coil
MRI safety
peripheral nerve stimulation

Journal

Magnetic Resonance in Medicine cover
Magnetic Resonance in Medicine
IF:
3
Papers:
1.2W
Citations:
3.1W

Organization

I
ihsan dogramaci bilkent university
Scholars:
3.6K
Papers: 3.5K
Citations: 8