arrow
Return

Multifocal multiphoton endoscope

delete2012-04-11
delete41
PRE
AI
D
David R. Rivera *
C
Christopher M. Brown
D
Dimitre G. Ouzounov
W
Watt W. Webb
C
Chris Xu
DOI:10.1364/OL.37.001349delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
We report a miniaturized resonant/non-resonant multi-fiber raster scanner that is paired with a gradient-index lens assembly to achieve a compact and flexible multifocal multiphoton endoscope capable of longitudinal parallel image acquisition. Multiphoton images are obtained simultaneously at three axial depths, separated by >= 4.8 mu m, by incorporating three axially offset double clad optical fibers into the miniaturized scanner. The fabricated endoscope has an outer diameter of 3 mm, a rigid length of 4 cm, and acquires images at 4 frames/s per focal plane, with lateral and axial resolutions for two-photon imaging of 0.8 and 10 mu m, respectively. (C) 2012 Optical Society of America
Keywords:
2-PHOTON MICROSCOPE
FLUORESCENCE
COMPACT
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

Optics Letters cover
Optics Letters
IF:
3.3
Papers:
4.0W
Citations:
7.6W

Organization

C
Cornell University
Scholars:
6.3W
Papers: 5.4W
Citations: 10.9W