PARTIAL FLIP ANGLE MR IMAGING

delete1987-02-01
delete92
PRE
AI
T
Timothy C. Mills
D
D A Ortendahl
N
NM Hylton
L
L E Crooks
J
Joseph W. Carlson
L
L. Kaufman
DOI:10.1148/radiology.162.2.3797669delete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En
Theoretical analysis predicts that performing magnetic resonance (MR) imaging with partial (< 90.degree.) flip angles can reduce imaging times two- to fourfold when lesions with elevated T1 values are being examined. This time savings occur because repetition time (TR) is reduced when imaging is performed with partial flips. Partial flip MR imaging can also improve signal-to-noise ratio (S/N) in fast body imaging. For this study, analytical tools were used to predict image contrast and S/N for short TR, partial flip sequences. Experimental implementation of the short TR, partial flip sequences that analytical work had predicted would be optimal supported the analytical predictions and demonstrated their validity. Partial flip MR imaging is applicable to reducing imaging time only when the ratio of signal differences to noise exceeds threshold values in conventional MR images. Partial flip sequences can be used to advantage in MR imaging of both the head and the body, and the observed effects are predictable through theoretical analysis.