Phase-encode reordering to minimize errors caused by motion
DOI:10.1002/mrm.1910350318.png)
Abstract
En 中文
A new method for suppressing the effects of motion in MR images by reordering the acquisition of k space has been developed. Existing reordering methods suffer from image blurring, The method presented here applies specifically to translation along the phase-encoding direction, in which case it reduces both ghosting and blurring. The method is conceptually similar to a linear phase shift of k space, except that it has the advantage of not corrupting stationary structures inadvertently. The method is intended for anatomic sites in which substantial translational motion occurs along the phase-encoding direction, such as the cranial-caudal motion of the liver and kidneys, The reordering method is motivated from an analysis of factors affecting the severity of motion artifacts. The theory behind the reordering method is described and validated experimentally by imaging a moving phantom.
Keywords:
motion artifacts
phase-encode reordering