arrow
Return

Phosphate Sensor Using Molybdenum

delete2016-06-24
delete43
PRE
AI
Y
Yinhua Li *
T
Taoqing Jiang
余晓冬 (Xiaodong Yu)
H
Huizhong Yang
DOI:10.1149/2.0161609jesdelete
deleteOriginal
deleteOriginal request for help
deleteShare
deleteSave
Abstract

Abstract

En 中文
Aphosphate ion-selective electrode (ISE) based on molybdenum was developed, which is based on the reaction between molybdenum and hydrogen phosphate ions (HPO42-) in alkaline condition. We found that the electrode showed a near Nernst characteristics with a slope of -26.9 +/- 0.5 mV/decade. Potential linear range is from 10(-1) M to 10(-5) M and detection limit is 1.9 x 10(-6) M in the pH of 8.5. The short response time for HPO42- ions indicates its fast response characteristic. The ISE can be used over a period of three months with good reproducibility and selectivity. The manufacture process of the ISE is easy and inexpensive. It can be applied to quantitative analysis of phosphate in water. (C) 2016 The Electrochemical Society. All rights reserved.
Keywords:
CARBON-PASTE ELECTRODE
INORGANIC-PHOSPHATE
IONIC LIQUID
COBALT FILM
MICROELECTRODE
FABRICATION
PHOSPHORUS
MEMBRANE
SILICATE
PYRUVATE
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

Journal of the Electrochemical Society cover
Journal of the Electrochemical Society
IF:
3.3
Papers:
3.3W
Citations:
9.4W

Organization

J
Jiangnan University
Scholars:
3.9W
Papers: 2.7W
Citations: 4.7W