Reduced aliasing artifacts using variable-density k-space sampling trajectories
DOI:10.1002/(SICI)1522-2594(200003)43:3<452::AID-MRM18>3.0.CO;2-B.png)
Abstract
En 中文
A variable-density k-space sampling method is proposed to reduce aliasing artifacts in MR images. Because most of the energy of an image is concentrated around the k-space center, aliasing artifacts will contain mostly low-frequency components if the k-space is uniformly undersampled. On the other hand, because the outer k-space region contains little energy, undersampling that region will not contribute severe aliasing artifacts. Therefore, a variable-density trajectory may sufficiently sample the central k-space region to reduce low-frequency aliasing artifacts and may undersample the outer k-space region to reduce scan time and to increase resolution. In this paper, the variable-density sampling method was implemented for both spiral imaging and two-dimensional Fourier transform (2DFT) imaging. Simulations, phantom images and in vivo cardiac images show that this method can significantly reduce the total energy of aliasing artifacts. In general, this method can be applied to all types of k-space sampling trajectories. (C) 2000 Wiley-Liss, Inc.
Keywords:
MRI
aliasing artifacts
spiral trajectory
variable-density sampling
Nyquist criterion
AI Summary
Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.
Journal
IF:
3
Papers:
1.2W
Citations:
3.1W
Organization
No organization information available
Cited Papers
Oncoids and groundwater calcrete in a continental siliciclastic succession in a fault-controlled basin (Early Permian, Northern Italy)
Facies
IF0