This study used Monte Carlo method to simulate particle movement inside a commercialized instrument called Subcutaneous Injection Site Simulator (SCISSOR). A series of parameters related to instrument, injection device, operation, formulation, as well as medium were thoroughly investigated. Also, several events that particles may occur in the subcutaneous (SC) space, including diffusion, binding, and aggregation, were implemented in our Monte Carlo based algorithms. The simulation result revealed that membrane area and position, diffusivity in the simulated SC medium, as well as injection position significantly affected release profile. Diffusivity in the injection volume could only influence release profile when this diffusivity was extremely low. Other factors, including initial injection shape, injection volume, and formulation concentration, had only minor impact on release profile. In addition, binding slowed down release, whereas aggregation reduced both total percentage of release and release rate. This study presented Monte Carlo method would potentially become a powerful tool to support multiple development activities related to experimental design, parameter sensitivity analysis, and control strategy development.
Keywords:
Monte Carlo simulation
Subcutaneous formulation
SCISSOR
Diffusion
Binding
Aggregation

Journal

International Journal of Pharmaceutics cover
International Journal of Pharmaceutics
IF:
5.2
Papers:
2.2W
Citations:
6.7W

Organization

U
University of Kansas
Scholars:
1.9W
Papers: 1.7W
Citations: 8.1K