This work presents the modelling, design, fabrication and test of the first micro Coriolis mass flow sensor fully fabricated in SU-8 by photolithography processes. The sensor consists of a channel with rectangular cross-section with inner opening of 100 mu m x 100 mu m and is actuated at resonance by Lorentz forces. Metal tracks for the actuation current are deposited on top of the chip. The chip has been tested over a flow range of 0-800 mg/min in the case of water and 0-650 mg/min for isopropyl alcohol (IPA) to confirm that the sensor measures true mass flow (at the maximum flow of the tested range, 0-800 mu l/min for both water and IPA, it results in a pressure drop of 0.98 bar for water and 1.97 bar for IPA). (C) 2016 Elsevier B.V. All rights reserved.
Keywords:
Flow sensor
Coriolis
Microfluidics
SU-8
MEMS
AI Summary

AI Summary

Key information extracted from the uploaded paper, including a brief overview, abstract, background, key highlights, visual analysis, and future outlook.

Journal

Sensors and Actuators B-Chemical cover
Sensors and Actuators B-Chemical
IF:
7.7
Papers:
3.3W
Citations:
12.6W

Organization

C
ciber - centro de investigacion biomedica en red
Scholars:
3.5W
Papers: 2.5W
Citations: 45
I
Instituto de Salud Carlos III
Scholars:
1.4W
Papers: 8.7K
Citations: 2.3W